Preparative Requirements for Volatile, Semi-volite, and Gas Chromatography Quality Control

Effective Date: 07/2008
Point of Contact: Quality Engineer for Regulatory Programs

Quality Control Requirements Frequency Frequency Failure Action
Method blank 1 per batch < EQL (See Organic Quality Control exhibit) Investigate against samples and client DQRs. Correct. Re-prepare as required.
BS

or

LCS
1 per batch (Blank, spiked with surrogate)

1 per batch
See Organic Quality Control exhibit Investigate against samples and client DQRs. Correct. Re-prepare and analyze as required.
MS and MSD
(Accuracy)
(Precision)
1 set per batch See Organic Quality Control exhibit Evaluate. If lab error, re-prepare and analyze. If matrix driven, evaluate against DQRs, notify client if still unacceptable, discuss in narrative.
Surrogate Each sample, QC sample, standard, and blank. See Organic Quality Control exhibit Evaluate. If lab error, re-prepare and analyze. If matrix driven, evaluate against DQRs, notify client if still unacceptable, discuss in narrative.
Total Carbon, Total Inorganic Carbon, and Total Organic Compound
Duplicate 1 per batch <20% RPD when >10 times IDL Evaluate. If lab error, re-prepare and analyze. If matrix driven, evaluate against DQRs, notify client if still unacceptable, discuss in narrative.
MS 1 per batch Recommended
75% to 125%
Evaluate. If lab error, re-prepare and analyze. If matrix driven, evaluate against DQRs, notify client if still unacceptable, discuss in narrative.
Blank Spike or

Laboratory Control Sample
1 per batch 80% to 120%

Vendor or Statistical
Investigate, evaluate against DQRs, correct, re-prepare and analyze if applicable.
Method blank 1 per batch <EQL Investigate, evaluate against DQRs, correct, re-prepare and analyze if applicable.

BS = Blank Spike

EQL = Estimated quantitation limit

MS = Matrix spike

MSD = Matrix spike duplicate

RPD = relative percent difference.