Contents

Physical Testing Quality Control (Differential Scanning Calorimetry, Thermogravimetric Analysis, and Differential Thermal Analysis/Thermogravimetric Analysis)

Effective Date: 11/01
Point of Contact: Quality Engineer for Regulatory Programs

Quality Control Requirements Frequency Acceptance Criteria Failure Action
Differential Scanning Calorimetry
Continuing calibration verification (performance check) temperature characteristics Each time the temperature ramp rate is changed Standard(s) which is within the range of analysis and thermogravi-metric analysis standard (material of known thermal properties) shall conform to known physical constants. Investigate. Correct and restart. Recalibrate if necessary.
Thermogravimetric Analysis, Differential Thermal Analysis/Thermogravimetric Analysis
Continuing calibration verification (performance check) mass/weight characteristics Periodically, not to exceed quarterly Check balance performance using material of known mass or of known mass loss as a function of temper-a-ture properties Investigate. Correct. Restart. Recalibrate if necessary.
Rheology
Continuing calibration verification Monthly or within 30 days of measurement ± 15% at viscosities <10CP ± 10% at viscosities >10CP Repair or Replace
Particle Size Analysis
Control Check Standard Daily ± 10% of expected value Rerun standard. If fails again, then investigate and correct.