Contents

Calibration Requirements for Inorganic Analyses

Effective Date: 08/06
Point of Contact: Quality Engineer for Regulatory Programs

Instrument or Test Calibration Criteria Frequency Acceptance Criteria Failure Action
Inductively coupled plasma spectrometer Profile Blank and one concentration per analyte of interest Before each new analytical run(a) Meet manufacturer's recommendation for peak centroid and full width at half max. A calibration verification standard within ±10% of true value confirms acceptability of calibration Investigate. Correct. Recalibrate, as required.
ICP/mass spectrometer (MS) Tune
Check mass calibration
Calibrate using at least a blank and one standard according to manufacturer's recommendations.
Analyze four times.
Before each analytical run.
RSD ≤5%
Mass calibration must be less than 0.1 amu of actual value. Resolution must be less than 0.9 amu full widths at 10% peak height. Calibration verification within ±10% of true.
Investigate. Correct. Adjust mass calibration to correct value.
Investigate. Correct. Recalibrate.
Atomic absorption: graphite furnace or flame Blank and three concentrations per analyte of interest, working range Before each new analytical run(a) Calibration curve must have correlation coefficient > 0.995. A mid-range calibration verification standard within ±10% of true value confirms acceptability of calibration Investigate. Correct. Recalibrate, as required.
Cold Vapor Mercury Blank and 4 standards Before each new analytical run Calibration curve must have a Correlation of Coefficient > 0.995. A Midrange calibration verification standard within ±20%. Investigate. Correct. Recalibrate, as required.
Cyanide (spectrophotometric) Blank and three concentrations (undistilled) to define the working range Before each new analytical run(a) Calibration curve must have a correlation coefficient > 0.995. A distilled mid-range calibration verification standard within ±15% of true value confirms acceptability of calibration Investigate. Correct. Recalibrate, as required.
Ion chromatography Blank and three concentrations to define the working range After major maintenance or upon failure of the calibration verification Calibration curve must have a correlation coefficient >0.995. A mid-range calibration verification standard within ±10% of true value confirms acceptability of calibration Investigate. Correct and recalibrate, as required.
pH Two-point calibration; 3 pH units or more apart calibration range representative of sample results. Note: For corrosivity characterization, one buffer should be a pH of 2 for acidic wastes and pH of 12 for caustic wastes. Before each new analytical run(a) For dial or slope calibration, analysis of calibration standards must measure within 0.05 pH units of true value A calibration verification standard within 0.1 pH units of true value confirms acceptability of calibration Investigate. Correct. Recalibrate, as required.
ISE
Working curve
Minimum of 3 concentrations, with one near the MDL, to define the working range Before each new analytical run(a) A mid-range calibration verification standard within ±10% of true value confirms acceptability of calibration Investigate. Correct. Recalibrate, as required.
ISE Known Addition, etc. Follow manufacturer's recommendations. All samples, standards and blanks Calibration verification within ± 3 standard deviation of the historical mean Investigate. Correct. Recalibrate, as required.
Colorimetric analyses (excluding cyanide) Blank and a minimum of 3 concentrations, with one near the MDL, to define the working range Before each new analytical run(a) A mid-range calibration verification standard within ±10% of true value confirms acceptability of calibration Investigate. Correct. Recalibrate, as required.
Titrimetric N/A N/A Standardize titrants before use. Evaluate results based on QC performance. Investigate. Correct. Recalibrate, as required.

(a) Analytical run is defined as sequence of analyses within a continuous time period.

ISE = Ion selective electrode

MDL = Method detection limit.