Contents

Analytical Requirements for Total Carbon, Total Inorganic Carbon, and Total Organic Carbon Quality Control

Effective Date: 07/2008
Point of Contact: Quality Engineer for Regulatory Programs

Quality Control Requirements Frequency Acceptance Criteria Failure Action
Continuing calibration verification Every 15 samples and at the end of the run 85% to 115% Rerun previous 10 samples
Continuing calibration Blank Periodically, following continuing calibration verification Evaluate for carry-over and contamination Investigate. Correct. Reanalyze.