Contents

Inorganic Analytical Quality Control Requirements

Effective Date: 07/2008
Point of Contact: Quality Engineer for Regulatory Programs

Quality Control Requirements Frequency Acceptance Criteria Failure Action
Inductively Coupled Plasma Spectrometer (ICP) and Flame Atomic Absorption (FLAA)
ICV Immediately after calibration (typically mid-range) 90% to 110% Investigate failure for analytes of interest. Recalibrate for analytes of interest.
ICB After ICV <EQL Investigate failure for analytes of interest. Recalibrate for analytes of interest.
CCV After every 10 samples and at the end of analytical run 90% to 110% Investigate failure for analytes of interest. Recalibrate for analytes of interest. Reanalyze samples, as appropriate.
CCB After every CCV <EQL Investigate failure for analytes of interest. Recalibrate for analytes of interest. Reanalyze samples, as appropriate.
LLS At the beginning of the analysis (but not before the ICV or ICB) Recommended:
75% to 125%
Investigate. Discuss in narrative.
Interference check standard (ICP only) After the ICV, after the ICB and at the end of the run 80% to 120% Investigate. Reanalyze all samples.
Serial dilution 1 per batch as required or needed <10%, when analyte >10 times IDL after 5-fold dilution Investigate. Re-prepare/Re-analyze as necessary. Discuss in narrative.
Post spike MS fails or new and/or unusual matrix is being analyzed Recommended: 75% to 125% Investigate. Re-prepare/Re-analyze as necessary. Discuss in narrative.
Inductively Coupled Plasma Spectrometry/Mass Spectrometry (ICP/MS)
ICV Immediately after calibration (typically mid-point region) 90% to 110% Investigate failure for analytes of interest. Correct. Recalibrate for analytes of interest.
CCV Every 10 samples and at the end of the run 90% to 110% Investigate failure for analytes of interest. Correct. Recalibrate for analytes of interest.
ICB After ICV < EQL Investigate failure for analytes of interest. Correct. Recalibrate for analytes of interest.
CCB After each CCV < EQL Investigate failure for analytes of interest. Correct. Recalibrate for analytes of interest.
Interference check standard After ICV and ICB and every 12 hours Monitor for interferences which will impact samples Investigate. Correct/Reanalyze/Flag.
Quality Control Requirements Frequency Acceptance Criteria Failure Action
Serial dilution 1 per batch of samples prepared or when internal standard criteria failure occurs ± 10% difference when analyte > 100 times IDL Investigate for analyst error. Discuss performance in narrative.
Post spike When MS fails or when new or unusual matrix is encountered 75% to 125% Investigate for analyst error. Discuss performance in narrative.
Internal standard Every sample, QC sample, blank, and standard 30% to 120% Perform serial dilution. Evaluate. Correct/Report.
Mercury, Cold Vapor Atomic Absorption (CVAA)
ICV(a) Immediately after calibration (typically mid-range) 80% to 120% Investigate. Correct. Recalibrate.
ICB After ICV < EQL Investigate. Correct. Recalibrate.
CCV After every 10 samples and at the end of analytical run 80% to 120% Investigate. Recalibrate. Reprepare and reanalyze samples, as appropriate.
CCB After each CCV < EQL Investigate. Recalibrate. Reprepare and reanalyze samples, as appropriate.
LLS After ICV and ICB Recommended: 75% to 125% Investigate for analyst error. Discuss in narrative.
Cyanide (Spectrophotometric)
ICV(a) Immediately after calibration (typically mid-range) 85% to 115% Investigate. Recalibrate.
ICB After ICV < EQL Investigate. Recalibrate.
CCV After every 10 samples and at end of analytical run 85% to 115% Investigate. Recalibrate. Reanalyze samples, as appropriate.
CCB After every CCV < EQL Investigate. Recalibrate. Reanalyze samples, as appropriate.
LLS After ICB Recommended: 75% to 125% Investigate. Discuss in narrative.
Quality Control Requirements Frequency Acceptance Criteria Failure Action
Graphite Furnace Atomic Absorption (GFAA)
ICV Immediately after calibration (typically mid-range) 90% to 110% Investigate. Recalibrate.
ICB After ICV < EQL Investigate. Recalibrate.
CCV After every 10 samples and at the end of analytical run (may be undistilled) 80% to 120% Investigate. Recalibrate. Reanalyze samples, as appropriate.
CCB After every CCV < EQL Investigate. Recalibrate. Reanalyze samples, as appropriate.
LLS After ICB Recommended: 75% to 125% Investigate. Discuss in narrative.
Post Spike When MS fails 75% to 125% Investigate for source of error. Re-prepare/reanalyze as necessary. Discuss in narrative.
Ion Chromatography (IC)
ICV After initial calibration (typically mid-range) 90% to 110% Investigate. Recalibrate.
ICB After ICV < EQL Investigate. Recalibrate.
CCV At the beginning of the analytical sequence when initial calibration is not run, after every 10 samples, and at the end of analytical run 90% to 110% Investigate. Recalibrate. Reanalyze samples, as appropriate.
CCB After every CCV < EQL Investigate. Recalibrate. Reanalyze samples, as appropriate.
LLS After ICB Recommended: 75% to 125% Investigate for analyst error. Discuss in narrative.
pH
ICV Immediately after calibration ±0.1 pH unit Investigate. Recalibrate.
CCV After every 10 samples and at the end of analytical run ±0.1 pH unit Investigate. Recalibrate. Rerun all samples since last valid CCV.
Ion Specific Electrode and Colorimetric/Spectrophotometric Methods
ICV Immediately after calibration for working curve technique, prior to analysis of samples for all other techniques 90% to 110% Investigate. Recalibrate.
ICB After ICV < EQL Investigate. Recalibrate.
CCV After every 10 samples and at end of the run (not applicable to titration methods) Based on long term statistical performance Rerun all samples since last valid CCV.
CCB After each CCV < EQL Rerun all samples since last valid CCB.
Titrimetric Methods (e.g., Ammonia, Hexavalent Chromium)
ICB 1 per batch < EQL Investigate. Correct. Reanalyze.
Titrant Verification 1 per batch 80% to 120% Investigate failures. Correct. Re-standardize titrant and reanalyze samples.

(a)If the ICV is representative of the sample matrix and prepared with the samples, then the ICV may be used as both the BS and ICV.
AS = Analytical spike
CCB = Continuing calibration blank
CCV = Continuing calibration verification
ICB = Initial calibration blank
ICV = Initial calibration verification
LLS = Low-level standard
MS = Matrix spike.