Minimum Calibration Requirements for Externally Calibrated Equipment

Effective Date: 11/01
Point of Contact: Quality Engineer for Regulatory Programs

Equipment or Technique Calibration Requirements Criteria Frequency Corrective Actions
Thermogravimetric analysis Calibration performed by manufacturer Confirm acceptability using performance check (temperature and balance checks) Upon failure of performance check Recalibrate when performance cannot be re-established
Differential thermal analysis and thermal gravimetry Calibration performed by manufacturer Confirm acceptability using performance check (temperature and balance checks) Upon failure of performance check Recalibrate when performance cannot be re-established
Differential scanning calorimetry Calibration performed by manufacturer Confirm acceptability using performance check (temperature and enthalpy checks) Upon failure of performance check Recalibrate when performance cannot be reestablished
Particle size analysis Calibration performed by manufacturer Confirm acceptability using performance check Upon failure of performance check Recalibrate to manufacturer's recommendations
Rheology Calibration performed by manufacturer Confirm acceptability using performance check Upon failure of performance check Repair or replace
Analytical balances Calibration performed by external organization Confirm acceptability using performance check Annually or upon failure of performance check Recalibrate at specified interval or when performance cannot be reestablished
Digital Thermometers Certified by external organization Confirm acceptability using performance check Upon failure of performance check or annually Replace or repair
Thermometers (for critical temperature measurements) Certified by external organization prior to receipt Confirm acceptability using performance check Annually, at ice point Replace or recertify
Pre-calibrated glassware Calibration performed by manufacturer Manufacturer's specifications meet analytical needs Not applicable Replace when out of specifications or contaminated
Conductivity meter N/A Confirm acceptability using performance check Upon failure of performance check Repair or replace
Density meter N/A Confirm acceptability using performance check Upon failure of performance check Repair or replace
Manometer Calibration performed by manufacturer Confirm acceptability using performance check Annually Replace or have externally recertified